Power Instrument Sterilization Chart

MicroAire Sterilization Parameters for Power Instruments and Accessories

Power Instruments and Accessories Gravity-Displacement Steam Dynamic Air Removal (pre-vac)
Instrument Minimum Time & Temperature Min. Heated Dry Time Instrument Minimum Time & Temperature Min. Heated Dry Time
Large Bone Power Series: 6000, 7000, Pulse Lavage Single Instrument 30 min. full cycle @ 132–135°C (270–275°F) 8 min.* Single Instrument 4 min. full cycle @ 132–135°C (270–275°F) 8 min.*
Sterilization Tray 45 min. full cycle @ 132–135°C (270–275°F) 20 min.* Sterilization Tray 4 min. full cycle @ 132–135°C (270–275°F) 10 min.*
Small Bone Power Series: 1000, 2000, 5000 Single Instrument 30 min. full cycle @ 132–135°C (270–275°F) 8 min.* Single Instrument 4 min. full cycle @ 132–135°C (270–275°F) 8 min.*
Sterilization Tray 35 min. full cycle @ 132–135°C (270–275°F) 60 min.* Sterilization Tray 4 min. full cycle @ 132–135°C (270–275°F) 45 min.*
NiMH Batteries REF: 7505-710, 6640-710 Unwrapped 3 min. full cycle @ 132–135°C (270–275°F) No drytime* Wrapped OR Unwrapped 4 min. full cycle @ 132–135°C (270–275°F) 1 min.*
Sterrad® 100 Sterilization 44 minutes total expose time Sterrad® NX Sterilization 28 minutes full cycle

*Additional times listed in corresponding product Instructions for Use (IFU).
CH-PWRSTER Rev A